Specimen Conditioning

Print this Topic Previous pageReturn to chapter overviewNext page

 

 

Test specimens should be vacuum saturated with 0.3M NaOH for approximately 18 hours, following the procedure described in ASTM 1202-97: 9. A companion sample is saturated in the same container to be used for pore solution extraction (see Pore Solution Extraction Test Procedure). The saturation procedure is summarized as follows: immerse the specimens in the 0.3M NaOH solution contained in the vacuum container. Turn on the vacuum pump. When the pressure gauge shows maximum vacuum pressure (less than 1 mm Hg -133 Pa), keep the pump running for about 2 hours. With all valves closed, turn off the pump and maintain vacuum conditions for 18 hours. Open the air valve to release the pressure.

 

 

 